Emergency Lid release

Mechanical Emergency Door release

During a power failure, you will not be able to open the Centrifuge lid via the Display button
Mechanical override is provided to allow sample recovery
However this is for Emergency use only
This should Not be used for normal use.
Warning
The rotor can still be spinning at a high speed, if touched it can cause serious injuries.
Always wait till the rotor has stopped.
Proceed as follows
• Make sure the Rotor has stopped (view via the port in the Lid)
• Once stopped turn off the power via the inlet Switch
• Pull the Centrifuge slightly over the bench (as shown in (a)
• You will notice one (b) or some models two plastic plug (c). Plug
• Pop open with a screwdriver (tool) and pull downwards .
• The lid will pop open You can now retrieve your samples
c
After replace the Plastic plugs back into their original position.
Reconnect the centrifuge power line once the mains power has been restored.
Check all is working correctly by closing the lid, wait 5 seconds then press lid open.
If not working refit or Contact your Distributor