Calibration is used to check the potentiostat's health, and to "zero" many of the measurement circuits to
your laboratory environment. A warning does not necessarily indicate a critical failure, and Gamry can
use calibration information to determine the source of the warning.
Double-check the following:
•
Confirm all connections for the 200 Ω Calibration Cell.
•
Floating-ground cable is connected to calibration shield.
If possible, the Chassis Ground is attached to a known, good earth ground.
•
Click the Retry button, and the rest of the calibration restarts. Click the Ignore button for any other
calibration warnings that may appear, and continue to Step 3.
After a failed calibration attempt:
Find Calibration Results PC6-#####.txt in your My Gamry Data folder.*
•
•
E-mail the file, along with complete contact information, to [email protected]
* ##### is the serial number of your potentiostat.
Did you receive a
C
W
ALIBRATION
ARNING?
A F
AILED
C
ALIBRATION