Program Menu (continued)
Choose three buffers (cal 1, cal 2, cal 3)
from the set for calibration. For ease of use,
set the buffers in the order that you will use
them.
cal 1
cal 2
For example: Set 1 (factory settings)
cal 3
cal 1=7.000, cal 2=4.000 and cal 3=10.010
1.679 not selected.
These can be rearranged in any order
using
and
value and move on.
read
Press
to exit the Program Menu.
Concentration mode
c = 1
You can select one of four preset slope
values for 1 point calibrations (c = 1, c = 2,
c = 3, c = 4). If you are not sure of your
electrode type, or wish to do 2- (3, 4, 5)
point calibrations, do not select a slope as
the 315 will calculate the slope and use this
value rather than the factory setting.
c = 1 -59.2 mV/decade, e.g. for BF
Cl
–
, ClO
c = 2 -29.6 mV/decade, e.g. for CO
c = 3 +59.2 mV/decade, e.g. for Ag
+
K
, Na
c = 4 +29.6 mV/decade, e.g. for Ca
Cd
2+
, Cu
Select your required slope using
mode
press
on.
ISo.Pt.
You can enter the isopotential point (in mV)
of your electrode. The range is –99 to
+99 mV. The 315 has a preset isopotential
point of 0 mV. Change the isopotential
point using
the value and move on.
You can set five concentration standards to
values of your choice, the range is 0.001 to
19999. For ease of use, set the standards in
cal 1 to
the order that you will use them. Set the
cal 5
values using
the value and move on.
For example, to set 500
Press
or
10 will flash.
Press
or
(no thousands), then press mode to
confirm. The value changes to 000.0, with
the first 0 ('hundreds') flashing.
mode
. Press
to enter the
–
, F
–
, I
–
, NO
–
, NO
–
, SCN
4
2
3
+
+
, NH
4
2+
, Pb
2+
, WH
2+
to enter the value and move
mode
and
, press
to enter
mode
or
, press
to enter
; the display shows 10000, the
to change the value to 0000
Program Menu (continued)
–
–
, Br
,
4
–
2–
2–
, S
3
+
, H
+
,
2+
conc
,
and
,
read
Press
Buffer Tables
The 315 automatically corrects pH buffers for
temperature using the values shown in the table.
1.679 4.000 4.010 6.860 7.000 9.180 9.210 10.010
5°C 1.668 4.000 4.010 6.950 7.090 9.390 9.450 10.250
10°C 1.670 4.000 4.000 6.920 7.060 9.330 9.380 10.180
15°C 1.672 4.000 4.000 6.900 7.040 9.270 9.320 10.120
20°C 1.675 4.000 4.000 6.880 7.020 9.220 9.260 10.060
25°C 1.679 4.000 4.010 6.860 7.000 9.180 9.210 10.010
30°C 1.683 4.010 4.010 6.850 6.990 9.140 9.160 9.970
35°C 1.688 4.020 4.020 6.840 6.980 9.100 9.110 9.930
40°C 1.694 4.030 4.030 6.810 6.980 9.070 9.060 9.890
45°C 1.700 4.040 4.040 6.830 6.970 9.040 9.030 9.860
50°C 1.707 4.060 4.060 6.830 6.970 9.010 8.990 9.830
Press
or
to change the 'hundreds' digit
until the display shows 500.0, then press
mode
to confirm. The 'tens' digit will flash.
mode
Press
until the display reads 500.0
with the 'tenths' digit flashing, then press
mode
again to enter value.
For example, to set 0.5
Press
or
; the display shows 10000, the
10 will flash.
Press
or
to change the value to 0000
(no thousands), then press
confirm. The value changes to 000.0, with
the first 0 ('hundreds') flashing.
mode
Press
until the display reads 0.000,
with the 'tenths' digit flashing.
Press
or
to change the 'tenths' digit
until the display shows 0.500, then press
mode
to confirm. The 'hundredths' digit
will flash.
mode
Press
again, the display reads 0.500
with the 'thousandths' digit flashing, press
mode
again to enter value.
Select concentration units from the
following list: blank (no units), ppm, mg/g,
mol/L, mmol/L, g/L, mg/L, eq/L, meq/L, %,
ppt. The 315 is preset to blank (no units).
Use
or
to change units, press
to enter the units and move on
to exit the Program Menu.
mode
to
mode