avantor VWR VC100 사용 설명서 - 페이지 12

{카테고리_이름} avantor VWR VC100에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. avantor VWR VC100 19 페이지. Controller vacuum

Name
Set vacuum(mbar)
Hysteresis(mbar)
Time(min)
After setting the parameters, click "OK" key to return to the startup interface.
Note:
The larger the hysteresis value set, the larger the pressure fluctuation,
the smaller the hysteresis value set, the smaller the pressure fluctuation
and the more frequently the vacuum pump starts up and shuts down.
The recommended hysteresis value is set at 10mbar.
Note:
The sum of the set vacuum value and hysteresis value cannot exceed
(P0- 10) mbar and the set vacuum value is larger than the hysteresis value.
3. Program operation mode
Click "Vacuum control" button in the "Function" section to pop up the
program selection interface as shown in the following figure.
Set the vacuum value within a range of 1 - local atmospheric
pressure (P0)
Hysteresis value affects the pressure fluctuation range and
the ON/OFF frequency of vacuum pump.
Setting range 1-300mbar
Time can be set within a range of 0-5999min
Table 6
12
Function Introduction
Fig. 3.2.5