a&p instruments 2001 Series 사용 설명서 - 페이지 3

{카테고리_이름} a&p instruments 2001 Series에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. a&p instruments 2001 Series 10 페이지.

2. First measurement with the factory settings
There are only three steps to do:
2.1
Tare the balance to zero with the Tare button.
2.2
Place your sample material on the drying pan and spread it out to an equal
surface for a good drying Wait some seconds until the weight is shown
steady in the display.
2.3
Swing the radiator over the sample.
In this mode the sample will be dried until no more moisture evaporates and the
instrument will stop the drying process automatically.
The automatic parameters are set to " 8 mg weight loss, 20 sec interval, 1 min
start interval".
That means, the instrument will start the drying process ,wait 1 minute after the
start (start interval) until the weight loss is registered by the processor and it will
stop the drying process when in between 20 seconds ( measuring interval) less
than 8 mg weight loss happened.
Of course these parameters (and some more) can be adjusted to your demands.
3. Drying by timer or automatic function (Konsta Stop) ?
Most materials can be measured by using the automatic stop function. There are
only a few materials (e.g. sugar, jam), which could irritate the automatic function
and need the timer for a fixed drying time. Please use the timer only when your
material cannot be dried with correct reproducible results.
3