9. Failures shown in the display
ERR 2
ERR 3
ERR 4
ERR 5
ERR 10 - ERR 16
others
Other problems
Weight display runs continually and the small "g" does not appear.
The reason can be vibrations or too much air movement around the instrument or
there is dirt around the pan holder or this part is damaged or in contact with the
housing.
The instrument shows other irregular reactions, e.g. the radiator cannot be regulated
or the parameters are lost.
Switch off and on the instrument
Check if the correct type of instrument is shown (2011, 2031 or 2081)
If there is shown only "20_1", the instrument has lost its memory
Set the dimmer knob with the white line to the right side
Push the buttons "Set, Show and Change" together at the same time until a "beep"
sounds
Set the correct instruments type with the Tare button
Leave this mode with the Set button
Set all parameters with the Tare and Change button again.
The instrument should now work correct again.
If there are other problems or if you have any questions left, please contact
No sample on the pan
Overflow in ATRO range
No pan on the balance, pan holder
touches housing
Overflow sample weight
Place is not calm, pan holder is bent, dirt
around pan holder
Contact a& p service department