A-M Systems 3000 사용 설명서 - 페이지 7
{카테고리_이름} A-M Systems 3000에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. A-M Systems 3000 18 페이지. Ac/dc differential amplifier
can significantly reduce unwanted interference from the power source, it will cause
some distortion of the signal, especially in frequencies below 100 Hz. Therefore, the
Notch Filter should only be used if other noise reduction techniques such as proper
grounding and shielding are inadequate.
This rotary switch enables the user to select the upper boundary frequency
LOW PASS:
that the input signal begins to be cutoff. Signals above the cutoff frequency will be
attenuated by a factor of 100 (-40 dB) per decade increase in the input signal
frequency. For example, if the
will be attenuated by a factor of 100 while a 100 kHz signal will be attenuated by a
factor of 10,000. One of the uses of this filter is to reduce high-frequency noise that is
above the frequency content of the signal being recorded.
Head stage indicator
This LED indicates whether the unit is configured for use with (
HEAD STAGE:
lit) or without (
STAGE
for use with a head stage pg. 5).
Power Supply
: This toggle switch is the main power switch, controlling the DC power input to
POWER
the main circuit of the instrument. The LED next to the toggle switch is lit when the
instrument is
. This button does not control the power input to the remote AC
ON
Power Supply. Once the remote AC Power Supply is plugged in there is 30 Volts DC
to the unit. If this situation is undesirable please unplug the remote AC Power Supply
when the instrument is not in use.
A-M Systems
LOW PASS
off) a head stage (See configuring the Model 3000
HEAD STAGE
131 Business Park Loop, P.O. Box 850 Carlsborg, WA 98324
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website:
switch is set at
1 kHz
http://www.a-msystems.com
, then a 10 kHz signal
HEAD
4