Appendix 1:
Stage 1: Re‐seating the processor to the correct position: 
 
1. Turn off the SQA and disconnect the power supply cable.  
2. Remove the 4 screws on the rear panel using a Philips screw 
driver #2 (Fig.1) and open the SQA (Fig.2) 
3. Slightly press the center of the processor with an index finger 
to re‐seat it into the correct position. (Fig. 3). 
4. Close the SQA and replace the 4 Phillips screws on the rear 
Panel.  
5. Connect the power cable of the SQA‐V. 
6. Turn the SQA on and run the SELF TEST.  
7. If the SQA‐V passes, the repair process is complete.  
8. If the SQA‐V does not turn‐on or fails self‐test, go to stage 2. 
Stage 2: Replace the damaged processor: 
1. Turn off the SQA and disconnect the power supply cable.  
2. Release the 4 screws on the rear panel using a Philips screw 
driver #2 and open the SQA.
3. Remove the damaged processor using tweezers (see Fig.4).
4.    Replace the old processor with a new processor according to 
the following directions: 
Align the "dot" which is marked on the processor with the 
"Arrow" mark in the processor cavity (see Fig. 5‐6). 
5. Slightly press the center of the processor with an index finger 
to re‐seat it into the correct position (Fig. 7). 
6. Close the unit and screw back in the 4 Phillips screws on the 
rear Panel. 
7. Connect the power cable, turn on the SQA and verify that it 
passes the SELF‐TEST. 
 
File Name: Tech Bulletin_SQA_Blank Screen_12_OCT_10
Instructions for RE‐SEATING or REPLACING the SQA PROCESSOR 
  
 
www.mes-global.com
 
 
 
 
 
 
Fig. 1‐2: Open the SQA‐V 
 
 
 
 
 
 
Figure 3: Apply downward pressure to the processor
 
         Fig. 4: Remove the damaged processor
Dot mark on
the processor
Fig. 5‐6: Align the processor in the cavity correctly 
 
 
 
 
 
 
 
Figure 7: Re‐seat the processor
www.a-tech-global.com
 
 
 
 
Arrow mark on
the processor
 
 
Page 2 of 3