12. Disassembly of the DIN version
a. Secure the vise mounting tool into a bench vise with
the DIN adapter facing upward. Remove the protective
DIN cap (36). Using a 6mm hex key, unscrew the
DIN fitting (38) and remove the handwheel (39).
b. Using an 1/8" wooden dowel or similar sized tool,
push the filter (13) out of the DIN fitting. Remove
the filter o-ring (12).
c. Remove the o-ring (37) from the face of the DIN fitting.
13. Loosen the vise and remove the first stage. Remove
the vise mounting tool from HP port. Using a 4mm hex key,
remove all the port plugs (15 & 17) and o-rings (14 & 16).
14. Pull on the narrow end of the protective sheath (11) and
slide it off the body (10).
THIS ENDS DISASSEMBLY
Before starting reassembly, perform parts
cleaning and lubrication in accordance with
Procedure A: Cleaning and Lubricating
(p. 15).
REASSEMBLY PROCEDURE
NOTE:
Before performing any reassembly, it is
important to inspect all parts, both new and those
that are being reused, to ensure that every part
and component is perfectly clean and free of any
dust, corrosion, or blemishes. Before dressing
each o-ring with Christo-Lube®, check to ensure
it is clean, supple, and free of any blemish.
WARNING:
Use only genuine Aqua Lung parts,
sub-assemblies, and components whenever
assembling any Aqua Lung product. DO NOT
attempt to substitute an Aqua Lung part with
another manufacturer's, regardless of any
similarity in shape, size, or appearance.
Doing so may render the product unsafe, and
could result in serious injury or death.
1. Slide the protective sheath (11) on the first stage body
(10). Make sure all the ports and inlet fitting opening align
with the holes in the sheath.
. Install new, lubricated o-rings (14 & 16) on all the port
plugs (15 & 17). Install all the port plugs, except for one HP
port, and tighten them using a 4mm hex key.
3. Install a new, lubricated o-ring (8) onto the crown (9). Slide
the crown onto the seat extraction tool (p/n 109437) with the
sealing edge against the plastic handle. Insert the crown into
the body and press it into place. Use the blunt end of the seat
extraction tool to make sure the crown is properly seated.