4.4.
Storage
1. Do not discard the protective cap(s) that came with the sensor. If the sensor is removed from the
process for an extended period of time, thoroughly clean the sensor, put a piece of cotton ball with
few drops of water into the protective cap and replace it on the sensor. This keeps the junction from
drying out which causes slow response when put back into operation or causes permanent damage
to the sensor. Sensors should not be left in dry lines or empty tanks for extended periods.
2. Do not store the sensors in a dry or humid location. When storing, check the protective cap(s)
regularly to make sure the cotton ball remains moist. Improper storage of sensors voids the
warranty.