Abaxis VetScan HM5 빠른 참조 매뉴얼 - 페이지 8
{카테고리_이름} Abaxis VetScan HM5에 대한 빠른 참조 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Abaxis VetScan HM5 19 페이지. Hematology analyzer
Abaxis VetScan HM5에 대해서도 마찬가지입니다: 빠른 참조 매뉴얼 (17 페이지), 설치 매뉴얼 (1 페이지), 설치 지침 (2 페이지)
Changing the Reagent Pack
Materials needed:
HM5 Reagent Pack, HM5 Normal Control and latex or nitrile gloves.
If "Change Reagent Pack" is used when installing new reagents, the analyzer will warn the
user when there are fewer than 10 tests left in any given bottle. This feature will also track
the number of samples, number of blanks, install date and the volume remaining in each of
the bottles.
1
2
1. From the Home Screen, touch the
Maintenance icon.
2. Touch Change Reagent Pack.
3. Follow the detailed directions on the
screen for changing the reagents. Touch
Next to advance through the various steps.
4. Run Quality Control to ensure the Analyzer
is within calibration.