SPLT™ USER'S MANUAL
Liquid contaminants will be forcibly ejected from the SPLT purge tube. Provision for collecting the purged
liquids should be considered when installing the SPLT.
Excessively dirty or wet filter elements can adversely affect pressure control.
2.2

ELECTRICAL

In order to take advantage of the automatic purging feature of the PPC, the SPLT valve MUST be connected to
the PPC external driver #8.
Connect like this if drivers connector with attached cable was included with SPLT:
Connect drivers connector to DRIVERS connector on back panel of PPC. Attach the two wires to the
terminal block on the SPLT
Connect like this if you have SPLT without drivers connector:
Solder (2) 24 gauge wires of desired length to the 12-pin connector provided with the PPC. Connect
(1) wire to pin L and the other to pin B. Connect drivers connector to DRIVERS connector on back
panel of PPC. Attach the other end of these wires to the terminal block on the SPLT.
Polarity is NOT significant to the operation of the purge valve.