2.4 Start up

Once electrical requirements have been fulfilled, install the equipment in a clean
environment free from dust and/or corrosive fumes, far from direct sunlight and from any
directional lamps pointing to it.
Remove the sipper protector. Disengage the pump tubing by removing one of the ends.
Stretch the tubes and make sure it is not stuck due to the lack of continuous use.
Place the tubes on the rotor once again.
1-Connect the instrument to the main line with power chord. Do not change the plug nor
interrupt grounding connections.
2-Install the equipment on a vibration free area, leaving a space of at least 10 cm behind the
instrument.
3-Place a thermal paper roll in the printer (see section 6.4) and connect external keyboard to
instrument.
4-Turn the instrument ON by using the switch on rear panel (see graphic 1.2). The display
will read: the instrument identification.
5-Press any key, and the following message will appear:
SYSTEM SET UP
PLEASE WAIT