3A HEALTH CARE ASPEED PROFESSIONAL 사용 설명서 - 페이지 6
{카테고리_이름} 3A HEALTH CARE ASPEED PROFESSIONAL에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. 3A HEALTH CARE ASPEED PROFESSIONAL 9 페이지.
ASPEED PROFESSIONAL does not need maintenance and/or lubrication, although it is required to perform some simple checks before each use:
• Check the integrity of the shell and the power supply cable.
• Using a finger, close the suction connector and verify that the vacuum level reaches 0.80 ÷ 0.85 bar approx.
• Verify that there is no disturbing noise that may be the symptom of a malfunction.
• Check that the cage is correctly positioned inside its housing. It must be aligned with the suction hole of the lid, so the valve of the float can
obstruct it when the sucked liquid is too much against the maximum quantity that can be assimilated by the vessel (figure 2a).
• Check that the float is mounted in the right position and that it is free to slide inside the cage (dirt and deposits can obstruct its movement).
PROBLEMS
Excessive noise
The unit turns on but it does not suck
The vacuum value cannot be adjusted
The protection fuse is activated any time
the device is turned on
The vacuum gauge does not work
Note: if you experience faults or malfunctioning problems different to those listed above, always and exclusively contact authorised
assistance centres.
Piston electric-compressor, with thermal protection.
Protection against electric shock
Fuse
Supply voltage
Adjustable vacuum level
Air flow
Dimensions
Weight
Noise level
Operation cycle
Class of risk according to the 93/42/EEC directive
Operating Temperature/Humidity
Storage and Transport Temperature/Humidity
Operating-storage air pressure
PERIODICAL CHECK FOR THE SAFETY OF THE DEVICE
PROBLEMS, CAUSES AND SOLUTIONS
POSSIBLE CAUSES
Damaged pump or obstructions in the internal
suction duct
- Damaged pump
- Vacuum regulator fully open. Connection
tubes disconnected and/or badly connected,
broken connection tubes. Container not in a
vertical position, full, or defective overflow
valve. Possible blockage of the hydraulic cir-
cuit inside the unit.
Damage to the internal hydraulic circuit or
obstruction of the connection tubes to the
aspiration unit
Pump damaged or in short circuit
Liquid penetrating the pneumatic circuit.
TECHNICAL SPECIFICATIONS
Single pump version
Low flow, high vacuum device
230V ~ 50Hz 150VA
0 ÷ -0,85 bar (85 kPa)
approx. 15 l/min.
330 x 185 x 210(H) mm
3,5 kg approx.
approx. 53 dBA
+10° C to +40° C/10% to 95% RH
–25° C to +70° C / 10% to 95% RH
min. 690 hPa - max 1060 hPa
11
SOLUTIONS
Send to the customer care service
- Send to the customer care service.
- Check the position of the vacuum regula-
tor. Check the connections and the integri-
ty of the tubes. Position the container in a
vertical position, check the overflow valve
(blocked) and/or replace the silicon tubes.
Send to the customer care service
Send to the customer care service
Send to the customer care service
Double pump version
High flow, high vacuum device
Class I
T1, 6A - 250V
230V ~ 50/60Hz 100VA
approx. 22 l/min.
4,2 kg approx.
approx. 63 dBA
Continuous use
IIa