ABB AP102 사용 설명서 - 페이지 6

{카테고리_이름} ABB AP102에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. ABB AP102 16 페이지. Ph/redox (orp) cartridge sensors with water-wash option
ABB AP102에 대해서도 마찬가지입니다: 매뉴얼 (16 페이지)

ABB AP102 사용 설명서

2 INSTALLATION

2.1 Typical Installation

A typical installation is shown in Fig. 2.1, with offset ported
flowcell arrangements illustrated in Fig. 2.3. A recommended
installation for samples which may have entrained air or are
subject to intermittent flow is shown in Fig. 2.2.
To prevent spillage
when cartridge is
removed.
Sample Flow
Fig. 2.1 Typical Installation
Fig. 2.2 Recommended Installation for Samples with Entrained Air or Intermittent Flow
4
To ensure that the cartridge
is always in solution.
Drain
Vent
A
Approximately 15°
Clamp (Optional)
IN
Fig. 2.3 Installation of Offset Ported Flowcell Assembly
To prevent spillage when
cartridge is removed.
B
Point B must be lower than point A
to ensure that the cartridge is always
in solution.
3
/
in NPT
8
Male Stud Couplings
(not supplied)
OUT
Recommended clamps
(not supplied)
Drain