AQUASOL AB-T-TDS 사용 설명서 - 페이지 3

{카테고리_이름} AQUASOL AB-T-TDS에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. AQUASOL AB-T-TDS 6 페이지. Table top meters

1.1 Product Packing
a) AB-T-TDS
– Meter
b) AME-CN-LG
– Conductivity Electrode – 1 Unit
c) AME-T-T10
– Temperature Electrode – 1 Unit
d) Supply Adapter
– 12V/1A
e) Operating Manual
2. Technical Parameters :
Model No AB-T-TDS
Parameter
Conductivity
Range
0.0 to 200.0
mS/cm
Accuracy
±1% FS
Resolution
0.1µS/cm
0.1mS/cm
ATC
0 to 60
Deg C
Power
12V / 1A
Size
240 × 170 × 80 mm
Weght
550 Gms
4
①--- Parameter mode icon
⑤---Temperature Value
②---Measuring value
⑥---Operation reminder
③---Time
⑦---Stable symbols
– 1 Unit
TDS
Salinity
Temp
0.0 to 100
0.0 to 100
0 to 100
PPT
PPT
Deg C
±1% FS
±1% FS
±0.2 + 1 Digit
0.1 PPM
0.1 PPM
0.1 Deg C
0.1 PPT
0.1 PPT
0 to 60
0 to 60
NA
Deg C
Deg C

3. Instrument Structure

3.1.1. Display
1
COND
25.0
7
MESSAGE :
Press SET to enter system parameter settings.
Press CAL to enter calibration mode.
6
Press SAVE to save current reading.
Long-time press SAVE to retrieve the stored data.
①--- Parameter mode icon
④--- Measurement unit
②--- Measuring value
⑤--- Temperature Value
③--- Time
⑥--- Operation reminder
3.2. Operation Key
The meter has 8 operation keys.
3.2.1. ON/OFF – Switch On/Off meter
3.2.2. CAL/LEFT – Calibration Key / Left Direction Key.
(a) When measuring state, press this button to enter the calibration
setting interface of the instrument.
(b) When in system time & manual temperature compensation status,
the key is a Left direction key.
3.2.3. SAVE/RIGHT – Save Record Key / Right direction key
(a) Under measurement status short press "SAVE" key can save
measurement data in the corresponding mode (the lower right
corner of screen will display serial number M+).
(b) Under measurement status, long press "SAVE" key can recall saved
data in the corresponding mode (RM and serial number will be
displayed in search operation).
(c)
Under system setting and manual temperature compensation mode,
it is used as the right direction key
3
2
2000/ 1/4
0 0 : 0 8
µS/cm
°C MTC
Press
or
to select the measurement mode.
⑦--- Stable symbols
5
4
5