Placing into Biomeme Thermocycler
Open the lid of the Biomeme thermocycler. Place your test strip, with the void
filling cap inserted, into a 3-well slot. Don't worry if the void filling cap feels
slightly loose. When the lid of the thermocycler is closed it will further secure
the caps into place.
It's important to make sure your Go-Strip is oriented correctly when placing
it into your thermocycler. Make sure the strip connections that are visible
through the void filling cap cutouts are facing the back of your thermocycler.
This will ensure sample 1 is always to the far left.
Navigate to the Biomeme App to begin your testing protocol. For further
instructional information about your thermocycler, contact Biomeme.