DiaTecne PulsePen WPP001-ETT 사용자 설명서 - 페이지 11

{카테고리_이름} DiaTecne PulsePen WPP001-ETT에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. DiaTecne PulsePen WPP001-ETT 16 페이지.

Power frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
Tab. 3: Electromagnetic Immunity
The PulsePen is suitable for use in the specified electromagnetic environment. The user of the PulsePen should
assure that it is used in an electromagnetic environment as described below.
Immunity
Test
Radiated RF
IEC 61000-4-3
Conducted RF
I.C. 61000-4-6
Interference may occur in the vicinity of equipment marked with the following symbol: &
30 A/m
IEC 60601-1-2
Test level
10V/m
80 MHz - 2.7 GHz
3V
150 KHz - 80 MHz
6V ISM bands and
amateur radio
150 KHz - 80 MHz
IEC 60601-1-2
Power frequency magnetic fields should be at
Test level
levels characteristic of a typical location in a
typical commercial or hospital environment.
Compliance
level
10V/m
Levels characteristic of a typical location in a
typical commercial or hospital environment.
Not Applicable
Levels characteristic of a typical location in a
typical commercial or hospital environment.
! 1 1
!
Electromagnetic Environment
User Manual PulsePen - V. 4.6