BioNano Genomics Saphyr 60325 시스템 안전 매뉴얼 - 페이지 8
{카테고리_이름} BioNano Genomics Saphyr 60325에 대한 시스템 안전 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. BioNano Genomics Saphyr 60325 9 페이지.
Ensure the instrument is installed on a clean, sturdy, level, bench without exposure to direct sunlight or heat
sources.
Do not set the instrument up where liquid or chemicals are used. Ensure that no liquids are spilled into the
instrument.
Do not operate the instrument in the presence of flammable gases or fumes.
Dimensions
Measurement
Height
Width
Depth
Weight
Environmental Considerations
This instrument is designed for indoor use only.
Element
Temperature
Humidity
Elevation
Ventilation
Air Quality
30253 Rev A, Saphyr System Safety Guide for Saphyr P/N 60325
Instrument
38 cm (15 in)
86 cm (34 in)
71 cm (28 in)
103 kg (227 lb)
Specification
Maintain a lab temperature of 19°C (66°F) to 25°C
(77°F).
Maintain a noncondensing relative humidity between
20–80%.
The instrument may only be operated at a location
where the elevation is 2,000m (6,500 ft) or less above
mean sea-level.
Provide at least 5 cm (2 in) of clearance behind the
instrument to allow sufficient ventilation and access
to instrument power connection.
Overhead clearance required for installation and
service is 93 cm (37 inch).
Operate the instrument in a Pollution Degree II
environment or better.
Page 8 of 9