Dickson KT8P0 시작하기 - 페이지 8

{카테고리_이름} Dickson KT8P0에 대한 시작하기을 온라인으로 검색하거나 PDF를 다운로드하세요. Dickson KT8P0 8 페이지. Remote sensing temperature recorder

Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?
Is the unit locked up? This can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the unit
may appear to be working, but the readings won't change, also the chart will not rotate.
Remove power and battery, and then re-power.
Why won't the unit power up?
Why won't the unit power up?
Why won't the unit power up?
Why won't the unit power up?
Remove the batteries and power adapter for a minute or two, this will reset the unit. The
unit should respond when the adapter is plugged in.

WARRANTY

WARRANTY
WARRANTY
WARRANTY
Dickson warrants that this line of instruments will be free from defects in material and
workmanship under normal use and service for a period of twelve months after delivery.
This warranty does not cover routine calibration and battery replacement.
For Specifications and Technical Support go to
FACTORY SERVICE AND RETURNS
FACTORY SERVICE AND RETURNS
FACTORY SERVICE AND RETURNS
FACTORY SERVICE AND RETURNS
Contact Customer Service (630.543.3747) for a Return Authorization (RA) Number before
returning any instrument. Please have the model number, serial number and a PO number ready
before calling.
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
Dickson ensures that the Replaceable Sensors / Instruments listed below were developed, tested
and validated together.
Replaceable Sensor Model: R400
Instrument Model: KT8P0, KT8P2, KT8P3
Operation
Operation
Operation
Operation: The Instruments and Replaceable Sensor models listed above were specifically
designed to work together. No other sensors will work with the Instrument models listed. No
other Instrument will work with the Replaceable Sensors listed. The Replaceable Sensor must be
connected to the Instrument for the Instrument to operate properly.
Calibration: Only the Replaceable Sensor is calibrated. The sensor, and all calibration defaults
Calibration
Calibration
Calibration
and adjustments are stored on the Replaceable Sensor. Accuracy is strictly controlled by the
sensor – no adjustments are made by the Instrument. Readings are sent from the Replaceable
Sensor to the Instrument for storage purposes only. Certificates of Calibration / NISTs are only
supplied for the Replaceable Sensor for this reason.
Recalibration: When a Replaceable Sensor is due for recalibration the existing Replaceable Sensor
Recalibration
Recalibration
Recalibration
can be replaced with a new Replaceable Sensor thus eliminating the need to return the entire
instrument for recalibration. Replaceable Sensors can be returned for recalibration as well.
Dickson KT8P Manual V6 52418
www.DicksonData.com
8