Dickson TH8P5 매뉴얼 - 페이지 4
{카테고리_이름} Dickson TH8P5에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Dickson TH8P5 8 페이지. Temperature & humidity chart recorder, 8” (203mm)
Dickson TH8P5에 대해서도 마찬가지입니다: 사용자 설명서 (9 페이지)
Up = On
Down = Off
Switches 1 - 6
For Model
7 day
All Models
24 hr
All Models
31 day
TH8P2/3/5
RH
All Models
Dew Point
All Models
-20°F - 120°F
All Models
40°F - 110°F
TH8P2/3/5
0°F - 200°F
TH8P3/5
-20°C - 50°C
All Models
5°C - 40°C
TH8P2/3/5
Remember to install the correct chart to match the corresponding switch setting
Recording Time
All the TH8 recorders feature user selectable recording times.
24 hr, all models
7 day, all models
31 day (TH8P2/3/5)
Temperature Ranges
-20°F - 120°F
40°F - 110°F
-20°C - 50°C
5°C - 40°C
0°F - 200°F
H / Dew Point
Dip switch #6 allows you to select whether RH or dew point is recorded by the blue pen and shown on the display for the
models with displays. (The red pen always records temperature.)
RH
Dew Point
Note: When the blue pen is recording dew point, the reading indicates the temperature at which dew will form or frost point,
so you should use the temperature scale on the chart to read dew point.
1
2
3
4
D
D
D
U
U
U
D
D
D
D
D
U
U
D
U
U
#1 down
#2 up
#1 down
#2 down
#1 up
#2 up
#3 down
#4 down
#5 down
#3 down
#4 down
#5 up
#3 up
#4 down
#5 down
#3 up
#4 up
#5 down
#3 down
#4 up
#5 up
#6 down
#6 up
5
6
D
U
D
U
U
D
D