7 Commissioning
•
Switch the instrument power ON. The display will read "SE.BR". This is the 'Sensor Wire Break'
display condition.
7.1
Setting the Alarm Trip Points
• To activate the quick alarm settings screen press either the "▲" or "▼" buttons, while the
DS1200 is displaying the moisture level. The DS1200 will flash the "F.SET" message
followed by a selection menu for Alarm 1 or 2 relay. Scroll to the required alarm and press
the "OK" button. The DS1200 then displays the message "SETP" momentarily before
displaying the current set point in % of scale.
Note: - that on some Models, the Hysteresis is also settable.
Power Up
-24.7
▲or ▼
F.SET
Select Alarm Channel
"REL 1"
REL1
"REL 2"
using ▲and ▼
OK
SETP
Set Alarm
Activation Point
50.5
using ▲ and ▼
as % of scale
Page 7
OK
Settings:
▲
Increase set point
▼
Decrease set point
OK
Save & exit the menu