PSS - Portable Sample System - User Manual
3.2 Now turn on the SADPmini2 (/Ex). Fig 14. and ensure that the display indicates that the
measuring sensor is dry. (Typically < -70°C dewpoint).
If not, refer to the instrument user manual.
Fig 14.
Press ON for 2 seconds
3.3 After purging for a minimum of 2 minutes, place a finger over the VENT on the Flow
Indicator until the desiccant head on the instrument is fully raised as shown in
Fig 15.
Fig 15.
Place finger over VENT tube
to raise the desiccant head and expose
the sensor to the gas
Pulling up the head will draw moisture from ambient air in the
work place into the head chamber where the sensor is exposed,
this will result in a contaminated sample and give
a false measurement. Fig 16.
The sample flow must always flow in one direction from the gas
source and pass the sensor in the head for an accurate result.
Fig 16.
WARNING
12
Do Not Pull Head Up