PSS - Portable Sample System - User Manual
7 Cleaning the PSS
The PSS instrument must only be cleaned with a soft cloth dampened with mild
soapy water only.
Do not use solvents, silicones or abrasives to clean the system.
8 Gases to Avoid
Most gases can be moisture tested with no calibration necessary between
sampling different gases. However, due to their corrosive or other properties not
all gases are compatible with this instrument. The table below details the gases
that should be avoided. If in doubt contact your authorised distributor.
Gas and Chemical Symbol
Ammonia (NH
Mercury (Hg)
Ozone (O
)
3
Chlorine (Cl
)
2
Hydrogn Chloride (HCI)
Sulphur Dioxide (SO
9 Specifications
Chassis:
Powder coated 304 stainless steel
Internal Pipework:
316 stainless steel
Internal Fittings:
316 stainless steel
Quick Connection (inlet) Swagelok®:
maximum connected pressure 2900PSI (200barg)
maximum un-connected (uncoupled) pressure 250PSI (17barg)
Flow Indicator:
0 - 20 litres
Carrying Bag:
Antistatic PVC foam faux leather with rubber feet.
Filter Unit (when fitted):
High grade stainless steel with removeable housing for access to
)
3
)
2
DO NOT
Sample Suitability
Not Permitted
Not Permitted
Not Permitted
Not Permitted
Not Permitted
Permitted when moisture content is
less than 100ppm(v)