Error Codes

Error codes are shown on the main display for approx. 2 seconds. The program then returns automatically to the previous mode.
Display/Problem
No segments appear
on the display
oL
UL
disErr
calErr
appErr
PrtErr
Max. weighing capacity is less
than indicated under "Specifications"
The weight readout is obviously wrong
If any other errors occur, contact your local Acculab customer service center.
14
Cause
No power available
The AC adapter is not plugged in
Battery is drained
The load exceeds the balance/scale
capacity
Weighing pan not in place
Something is touching
the weighing pan
Display overflow:
Value cannot be shown
on the display
Calibration parameter not met; e.g.:
– Balance/scale not zeroed
– Balance/scale is loaded
Weight is too light or
there is no sample on the balance/scale
with application in use
Data interface for
printing is blocked
The balance/scale was switched on
without the weighing pan in place
The balance/scale was not calibrated/
adjusted before weighing
Balance/scale not zeroed
Solution
Check the power supply
Plug in the AC adapter
Replace battery; recharge battery using
external charger
Unload the balance/scale
Place the weighing pan
on the balance/scale
Move the object that is
touching the weighing pan
Reduce load on the balance/scale
Calibrate only when zero is displayed
Press (ZERO) to tare the balance/scale
Unload the balance/scale
Increase the weight on the balance/scale
Contact the Acculab
customer service center
Place the weighing pan on the
balance/scale and press (ON/OFF)
Calibrate/adjust the balance/scale
Zero the balance/scale