Acculab SVI-10A 사용 설명서 - 페이지 4
{카테고리_이름} Acculab SVI-10A에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Acculab SVI-10A 10 페이지. Svi series professional precision industrial scale

ENGLISH
Key Functions / Description:
1. ON/OFF:
Press to turn scale ON or OFF.
2. ZERO/TARE:
Pushing ZERO/TARE key "zero's" display, pushing and holding ZERO/TARE key for more than 3
seconds accesses application menu.
3. "F" Key:
Function key for applications (Unit toggling, Counting & Totaling ) and menu scrolling.
4. CAL/PRINT:
Press and "hold" CAL/PRINT key (for approximately 3 seconds) and scale will automatically enter
calibration mode (displaying one of three weight values). Press and "release" CAL/PRINT key to print
displayed information when utilizing RS-232 interface.
5. LIGHT :
Turns on backlight during battery operation (Auto Off: One Minute).
Calibration:
Weight Calibration:
1
"CAL/PRINT" key for more than 3 seconds to enter the calibration mode.
Depending upon your Model a default weight will be displayed in "kg" (note: you have 3 weight values
2
to choose from)
In order to select additional cal-weight options, press and release the "F" key and the display will
3
toggle between selectable values based on model. (Please see Product Specifications)
After selection of the desired cal-weight option, press the "CAL/PRINT" key and release to confirm,
4
and the value will flash to begin calibration.
5
Place corresponding weight onto center of pan. If weight is accepted, the display will stop flashing and
show "selected kg" value in display.
2
ON / OFF
TARE / ZERO
1.
with the pan cleared and zero stability indicated in display, press and hold the
LIGHT
F
CAL / PRINT
2.
3.
4.
5.