Acculab VIC-212 사용 설명서 - 페이지 11
{카테고리_이름} Acculab VIC-212에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Acculab VIC-212 16 페이지. Electronic precision scales
Acculab VIC-212에 대해서도 마찬가지입니다: 사용 설명서 (16 페이지)
Visit www.balances.com for the best prices on Acculab scales
Specific Gravity
Purpose
Use this application program to determine the specific gravity of a sample. The result is displayed with one decimal place.
Beaker and wire not included with balance/scale.
Example: Determine the specific gravity of a solid.
Step
1. Select application program
2. Select Specific Gravity
3. Confirm setting
Symbol " " on the display: application is active
4. If necessary, zero the balance/scale
5. Start application program
6. Confirm the display, "airval"
7. Determine weight of the sample in air:
place sample on the balance/scale
8. Store weight-in-air value
9. Remove sample from balance/scale
10. Determine weight of sample in liquid:
connect wire and set up beaker
11. Confirm the display "Water"
12. Place sample in liquid
13. Store the weight-in-liquid value and view the result
14. Clear the display
15. Exit the Specific Gravity application
16. Reactivate Specific Gravity (if no other
application program has been selected)
17. Repeat procedure starting from Step 5.
www.balances.com is the original deep discount distributor
Key (or instruction)
(ZERO) > 2 sec
(F) repeatedly
(ENTER)
(ZERO)
(F)
(ENTER)
(ENTER)
(ENTER)
(ENTER)
(ZERO)
(ENTER) > 2 sec
(F)
Display
11