Olympus BX41-P 사용 설명서 - 페이지 22

{카테고리_이름} Olympus BX41-P에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BX41-P 36 페이지. Polarizing microscope
Olympus BX41-P에 대해서도 마찬가지입니다: 사용 설명서 (36 페이지)

Olympus BX41-P 사용 설명서
Fig. 24
@
Fig. 25
19
²
Cross lines
@
Conoscopic image
Note: During conoscopic observation, the center of the conoscopic image
@ and the intersection of the cross lines ² may not coincide.
However, employment of the universal infinity optical design ensures
that this disagreement will have no measurable effect during actual
observation.
However, if the disagreement is distracting, adjust in the same manner
as when using the U-CPA (for photomicrography), described in the
following paragraph.
When Using Polarizing Intermediate Attachment (U-CPA)
· Photomicrography
1. Using the tip of a mechanical pencial, etc., push out, in the direction of
the objective hole, the black rubber plugs inserted into the centering
holes of the primary hole where the 20X or 10X objective is inserted.
2. Pull out the Bertrand lens knob @ to the OUT (¦) position. (Fig. 25)
3. Adjust the position of the intermediate tube (Refer to page 7.)
4. Center the condenser. (Refer to page 18 in the BX51/52 instruction manual
or page 16 in the BX41 instruction manual.)
5. Push in the Bertrand lens knob @ to the IN ({) position for conoscopic
observation.