Olympus BX45 사용 설명서 - 페이지 5
{카테고리_이름} Olympus BX45에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BX45 40 페이지.
IMPORTANT
This microscope employs a UIS2/UIS (Universal Infinity System) optical design, and should be used only
with UIS2/UIS eyepieces, objectives and condensers for the BX2 series. (Some of the modules designed for
the BX series and objectives/eyepieces for the UIS series are also usable. For details, please consult Olympus
or the catalogues.) Less than optimum performance may result if inappropriate accessories are used.
SAFETY PRECAUTIONS
Fig. 1
²
Fig. 2
1. After the equipment has been used in an observation of a specimen that
is accompanied with a potential of infection, clean the parts coming in
contact with the specimen to prevent infection.
· Moving this product is accompanied with the risk of dropping the speci-
men. Be sure to remove the specimen before moving this product.
· In case the specimen is damaged by erroneous operation, promptly
take the infection prevention measures.
2. Install the microscope on a sturdy, level table or bench so as not to block
the air vents on the underside of the base.
@
Do not place the microscope on a flexible surface, as this could result in
blocking the air vents and cause overheating or a fire.
3. To prevent obstruction of the natural convection-based air cooling of the
microscope, make sure to leave at least 10 cm of free space between
walls or other objects, and the left, right and rear sides of the microscope
and the lamp socket when installing the microscope.
4. To avoid potential shock hazards and burns when replacing the light
bulb, set the main switch @ to "
cord from the wall outlet in advance. Whenever you replace the bulb
during use or right after use, allow the lamp socket ² and bulb to cool
before touching. (Figs 1 & 2)
Designated bulb
# The microscope also incorporate a fuse (this should be replaced by
the manufacturer or authorized agent).
5. Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the section
"PROPER SELECTION OF THE POWER SUPPLY CORD" at the end of
this instruction manual. If the proper power cord is not used, product
safety and performance cannot be guaranteed.
6. Always ensure that the grounding terminal of the microscope and that
of the wall outlet are properly connected. If the equipment is not
grounded/earthed, Olympus can no longer warrant the electrical safety
and performance of the equipment.
7. Never insert metallic objects into the air vents of the microscope frame
as this could result in electrical shock, personal injury and equipment
damage.
8. The power cord may be melt by the heat of lamp socket if the cord
comes in contact with it. Distribute the power cord at an enough dis-
tance from the lamp socket.
" (OFF) then disconnect the power
6V30WHAL (PHILIPS 5761)
BX45
1