Olympus BX51-P 사용 설명서 - 페이지 12

{카테고리_이름} Olympus BX51-P에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BX51-P 36 페이지. Polarizing microscope
Olympus BX51-P에 대해서도 마찬가지입니다: 정보 (6 페이지), 사용 설명서 (36 페이지)

Olympus BX51-P 사용 설명서
@
Fig. 9
³
@
Fig. 10
9
9
Mounting Rotatable Analyzer (U-AN360P-2)
1. Place a desired ND filter (30 mm diam.) in the empty hole @ as required.
2. Insert the rotatable analyzer (U-AN360P-2) ² as far as the click-stop
position. Then screw in the stopper knob ³. (Fig. 10)
When using the fixed analyzer U-ANT instead of the rotatable analyzer
U-AN360P-2, place the fixed analyzer in the test plate adapter U-TAD.
Position the fixed analyzer so that the positioning pin on the fixed
analyzer fits into the groove. The fixed analyzer will be held in place by
²
a magnet.
10
Mounting Test Plate Compensator
1. Loosen the test plate adapter (U-TAD) clamping knob @.
2. Insert the sensitive tint plate (U-TP530), quarter wave plate (U-TP137),
or any of the other compensators (6 types) into the test plate adapter,
and securely tighten the clamping knob.
(Fig. 9)
(Fig. 10)