FIg. 22
CAUTION
FIg. 23
(5) Clamp the stage with the stage rotation clamping knob.
10. This concludes centering of the optical axis for the 10X objective, which
will act as the reference objective. Now, center the other objectives with
the centerable nosepiece by engaging the objectives into the light path
one after another.
When Using Intermediate Attachment for conoscopic and
orthoscopic observation (U-CPA)
· Normal Observation
Make sure the 10X objective is inserted into the primary hole on
the centerable revolving nosepiece.
1. Pull out the Bertrand lens knob @ to the pulled-out (¦) position. (Fig. 23)
2. Center the condenser. (For details, see instructions of each microscope.)
3. Follow steps 9 and 10 in "When Using the Intermediate Attachment for
orthoscopic observation".
BX53/BX53M-P