Olympus BX60 사용 설명서 - 페이지 10
{카테고리_이름} Olympus BX60에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BX60 48 페이지. System microscope
Olympus BX60에 대해서도 마찬가지입니다: 빠른 시작 매뉴얼 (4 페이지)
-j
BX60
Fig. B
fig. 9
Installing therHalogen Bulb
(Rga. B,
(for both Transmitted/Reflected Ughts)
9.10)
O The appropriate bulb is a 12V. 100WHAL halogen bulb (PhBips 77241.
1. Futry loosen the lamp housing damping screw© on top of the lamp
housing cover with the provided AQen screwdriver.
2. Lift the lamp housing cover Q) upward to remove it (fig. 81
3. Turn the lamp socket Q) 90* in the direction indicated by the arrow.
4. Holdtng the bulb® with gloves or a piece of gauze, depress the bulb
damping levers ® and insert the bulb pins ® MV into the pin holes
ffl. Gently release the bub damping levers ® to their original positions
to secure the bulb. (figs. 9. 10)
* Do not touch the bulb with bare hands. If fingerprints are
accidentally left on the bulb, wipe the bulb with toft doth.
S. Slide the lamp housing cover onto the housing base from above.
Tighten the damping screw ® whle pressing downward on the cover.
(Fig. 3)
* Whenever you replace the bulb, first tum OFF the main switch
and wait for ths bulb and lamp socket to cool down.
fig. 10