1.
Connect RS-232C cable to the connector on the front panel of the control box BXC-CBB. (See the figure below.)
To connect to a PC, use a D-Sub9 pin (female)-D-Sub9 pin (female) straight-through cable. (If you use an
l
incorrect cable, the product may be damaged.)
The connector on the front panel of the control box BXC-CBB: D-sub9pin DCE assignment
l
Mating clamping screw: #4-40 UNC
l
The following table shows the settings for communication (Fixed value)
l
Baudrate
19200 [bps]
Data bit
8 [bits]
Parity
even
Stop bit
1 or 2 [bits] (switching the setting not required) *
Terminator
CR+LF
Flow control None
*The setting of the stop bits for communicating from the Host-PC to BXC-CBB. The setting of the stop bits for
communicating from BXC-CBB to the Host-PC is fixed to "2".
TIP:
If connecting to a device other than a PC, do so at your own risk. Refer to the following table for the connections.
No. Signal name I/O
2
RXD
OUT Transmitted data
3
TXD
IN
Received data
5
GND
—
Signal ground
7
RTS
IN
Request to send
8
CTS
OUT Clear to send
Pins not indicated are not internally connected and pins 7 and 8 are internally connected to each other.
3. Setup procedures
Function
24
3.12 Connecting cables