Olympus BXFM-ILH 사용 설명서 - 페이지 13

{카테고리_이름} Olympus BXFM-ILH에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BXFM-ILH 36 페이지. Focusing unit

Olympus BXFM-ILH 사용 설명서

4-2 Focusing Unit

@
Fig. 6
Fig. 7
1

Adjusting the Coarse Adjustment Knob Tension

²
}The rotation tension of the coarse adjustment knob should be adjusted
using the tension adjustment ring. The coarse adjustment knob tension
is preadjusted for easy use. However, if desired, you can change the
tension using the tension adjustment ring @. Turning the ring in the di-
rection of the arrow increases tension, and vice versa.
The tension is too low if the stage drops by itself of focus is quickly lost
after adjustment with the fine adjustment knob. In this case, turn the ring
in the direction of the arrow to increase tension.
}If you want to increase the tightness of the rotation tension adjustment
ring, insert the provided Allen screwdriver into the 4 holes ² on the ad-
justment ring and tighten.
2
Using the Rubber Fine Focus Adjustment Knob Cap
When the fine adjustment knob is fitted with the rubber knob cap, the
knob can be adjusted with a lighter force, a making fine focusing easier.
BXFM
(Fig. 6)
(Fig. 7)
10