Olympus BXFMA 사용 설명서 - 페이지 23

{카테고리_이름} Olympus BXFMA에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BXFMA 36 페이지. Motorized illumination system with motorized focusing unit

Olympus BXFMA 사용 설명서

SPECIFICATIONS

Motorized Illuminator with Motorized Focusing Unit BXFMA-F
Item
Optical system
Illuminator
Focusing unit
Revolving nosepiece
Observation tube
Installation on system
Power supply sources
Motorized control
Operating environment
· UIS2 (UIS) (Universal Infinity System) optical system
· Reflected Koehler illumination
· Tube magnification: 1X (FN 26.5: Super-widefield compatible)
· Observation method switching: Motorized BF (Brightfield)/DF (Darkfield) switching
· Aperture iris diaphragm: Motorized opening/closing (diameter 0.4 to 11.1 mm), centering
possible
· Field iris diaphragm: Manual lever opening/closing, centering possible
· Filter: Manual engagement/disengagement, up to 2 filters ( 32 mm filters)
· Available observation methods: 1 Reflected light brightfield observation, 2 Reflected light
· Light sources: 100W halogen bulb (standard),
100 W mercury burner or 75 xenon bulb
· Drive system: Motorized focusing using a stepping motor and ball screw.
· Resolution: 0.01 µm
· Highest speed: 5 mm/sec. (The factory setting of the BX-UCB control box is 3 mm/sec.)
· Full stroke range: 5 mm (2 mm down and 3 mm up from the standard focusing position)
(7 mm including the escape stroke)
· Motorized revolving nosepiece: U-D5BDREMC, U-D6REMC, U-P5REMC, U-D5BDREM or
U-D6REM
U-BI30-2
Widefield binocular
Widefield trinocular
tube
Interpupillary adjustment range: 50 mm to 76 mm
· Installable surfaces: 2 surfaces (left side panel, bottom panel)
· Installation method: 6 bolts
· Motorized controls: BX-UCB control box
· Light source: The halogen bulb is powered from the TH4 power supply unit or BX-UCB
control box.
· Possible using control units (U-HSTR2 hand switch and U-FH focus adjustment knob unit)
or PC (BX2-BSW software, commands).
· Indoor use.
· Altitude: Max. 2000 m
· Ambient temperature: 10° to 35°C (50° to 95°F)
· Maximum relative humidity: 15% to 85%
· Supply voltage fluctuations: ±10%
· Pollution degree: 2 (in accordance with IEC60664)
· Installation/Overvoltage category: II (in accordance with IEC60664)
Specification
darkfield observation
U-ETR3
U-TR30-2
Widefield erect
trinocular
FN 22
Tube tilting angle: Fixed
BXFMA
U-SWTR-3
Super-widefield
trinocular
FN 26.5
19