CK40M
3.

Absolute optical axis adjustment

3-1
Set the following jigs as illustrated below to adjust the absolute optical axis.
• Eyepiece------ KN0048: Standard eyepiece (with adapter-1)
• Objective ----- KN0041: Standard objective
• Other --------- BKN0003: Standard tube
C. REPAIR PROCEDURE
Standard objective
(KN0041)
C-3
Standard
eyepiece
(KN0048, with
Adapter-1)
Standard tube
(BKN0003)