CKX31/CKX41

1. Overview of optical adjustment

(1 ) Optical axis (CKX31)
<Fig.1>
1) After removing the stage, peel off the seals and remove
the screws (*1,*2).
Screws: 3PUTB2X4SA 1pc. (*1), CUKSK3X8SA 4pcs.(*2)
2) Separate the cover-L and cover-R.
Adjustment (B)
*6
*5
Adjustment (C)
Adjustment (A) : Loosen the screws(*4) and adjust the revolving axis by moving the right sleeve.
Adjustment (B ) : Loosen the screws(*5) and adjust the left /right optical axis by moving the left
Adjustment ( C) : Loosen the screws(*6) and adjust the absolute optical axis in the right sleeve
(2) Parfocality <CKX31>
Adjust the parfocality by inserting washer under
the lens frame-1.
(Check the parfocality in the left sleeve.)
D. REPAIR PROCEDURE
Adjustment (A)
Screw: CUK3X6SB 2pcs
sleeve. (the right sleeve is taken as a standard)
Screw: CUK3X6SB 2pcs. (*5)
by moving the binocular unit attached to the BI tube.
Screw: CUK3X6SA 4pcs. , Washer: KNW3SA 4pcs. (*6)
*When removing the BI tube, adjust the absolute optical axis by changing a
mounting position of BI tube.
Scerw : AB3X14SA ,Washer:BNW3SA 4pcs(*3)
Washer
(selection)
Lens frame-1
Cover-L
*1
*2
seal
*4
(*4)
.
<Right sleeve>
*2
Helicoid ass'y (B)
After checking it in the left sleeve, turn the
helioid to focus on the specimen in KN0041,
then loosen the screws(*1) ,only move the
diopter ring and fix it at the position where
moving range from focusing point is equal in
scewing -in /out directions.
Note: When using the screw(*2) for fixing eyepiece;
mount the diopter ring with (A) and (B) screw
hole positions aligned without the above
adjustment.
D-1
Cover-R
BI tube
<Fig.1>
<Fig.1>
Diopter ring (A)
(Positions of
screw holes)
*3