TROUBLESHOOTING GUIDE

Under certain conditions, performance of the microscope may be adversely affected by factors other than defects. If
problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after
checking the entire list, please contact Olympus.
Problem
1. The left and right fields of view do
not coincide.
2. Dirt or dust is visible in the field of
view.
3. Visibility of observed image is poor. Dirt/dust on the dustproof glass on the
4. Zooming causes defocusing of the
observed image.
5. The focus adjustment knob does
not rotate smoothly.
6. The microscope body drops spon-
taneously, causing the focusing to
be deviated during observation.
7. The right field of view or the moni-
tor image is cut off.
8. Zooming causes important defocusing
of the monitor image.
11
Cause
Interpupillary distance is adjusted im-
properly.
The parallax is not corrected.
The magnifications of the left and right
eyepieces are different.
Dirt/dust on specimen.
Dirt/dust on eyepieces
tip of the objective.
The eyepiece diopter is adjusted im-
properly.
The focus adjustment is inaccurate.
The rotation tension of the knob is set
too high.
The rotation tension of the knob is set
too low.
The light path selector lever is not set
properly to a stop position.
The confocality of the TV camera is ad-
justed improperly.
Remedy
Adjust it correctly.
Adjust it correctly.
Use the eyepieces with the same
magnification for the left and right by
replacing one of them.
Clean thoroughly.
Clean thoroughly.
Adjust it correctly.
Adjust the focus accurately.
Decrease the rotation tension to an op-
timum level.
Increase the rotation tension to an op-
timum level.
Push in or pull out the lever all the way
till the stop position.
Adjust it properly.
Page
6
6-7
3
2
2
6-7
5
6
6
9
10