³
ƒ
Fig. 12
Adjusting the Confocality of TV Camera
8
(SZ61TR)
}Ensuring the confocality between the observation image and the TV
camera monitor image makes it possible to maintain accurate focusing
of the observation image or monitor image even when the light path is
switched.
1. Adjust the diopter of the eyepiece (see pages 6 and 7) and bring the
specimen into focus.
2. Pull out the light path selector lever @ and set the magnification to the
²
minimum.
3. Slightly loosen the two confocality adjustment clamping screws ² and
@
C-mount clamping screw ³.
4. Hold the TV camera | and rotate the confocality ƒ to focus on the
monitor image.
After the adjustment, tighten the clamping screws ² and ³.
Applicable cameras (image pickup device size) and field numbers
}The TV camera light path incorporates a 0.5X lens.
Rotation of monitor image
The monitor image is rotated slightly because the right observation light
path, which is tilted slightly, is used as the TV light path.
The monitor image can be aligned with the observation image by rotating
the TV camera.
FN 22
(With 10X eyepiece)
SZ2
series
(Fig. 12)
2/3 in. image pickup device
(Diagonal length 22 mm)
1/2 in. image pickup device
(Diagonal length 16 mm)