Olympus U-CTB 사용 설명서 - 페이지 11

{카테고리_이름} Olympus U-CTB에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus U-CTB 24 페이지.

Olympus U-CTB 사용 설명서
5. Rotate the compensator knob and confirm that the measured portion on the center of the specimen is
dark as shown in Fig. 2-1 (Note). If it is not dark, rotate the stage by -90° and clamp.
(Note) Instead, a black interference may be observed with certain specimens (Fig. 2-2).
If the black fringe does not intersect the center of the field of view even after rotating
the stage 90°, the retardation of the specimen is outside the measurable range, and
measuring with the U-CBE or U-CTB is not possible. (See page 1)
U-CBE, U-CTB
8