Olympus U-CTB 사용 설명서 - 페이지 19

{카테고리_이름} Olympus U-CTB에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus U-CTB 24 페이지.

Olympus U-CTB 사용 설명서
U-CSE

4-4 Senarmont Compensator (U-CSE)

Measuring
} Always use interference filter (IF 546 or IF 550). If interference filter is not used,
measurement will not be possible.
# Due to its high detection sensitivity, use of the IF546 interference filter is recommended.
1
Adjusting of Analyzer Angle
# Perform the following adjustment with no specimen positioned on the stage.
1. Insert the U-CSE into the test plate adapter (U-TAD) or polarization attachment (U-PA) as far as it will go.
2. Fine adjust the polarizer of the condenser (U-POC) to obtain complete extinction.
3. Then fine adjust the rotatable analyzer (U-AN360P) to obtain complete extinction.
4. Repeat the steps 2 and 3 about 3 to 5 times.
5. At this point, read the analyzer's final angle. → θ
0
16