Olympus U-LH100G 사용 설명서 - 페이지 20
{카테고리_이름} Olympus U-LH100G에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus U-LH100G 42 페이지.
Problem
c) Mercury burner flickers or is dark.
d) Motorized controls do not work.
e) Hand switch or PC indicates that
shutter is released, but light is not
incident or image is cut off.
f ) Turret will not rotate (or stops in the
middle of movement).
g) Motorized revolving nosepiece will
not rotate.
Cause
It is soon after ignition.
Burner life has expired.
Burner is deviated from optical axis.
Connector are not correctly connected.
Shutter is moved due to external vibra-
tions.
Fluorescence mirror unit is slipping out
of turret.
Motorized revolving nosepiece connec-
tor is not correctly connected.
Remedy
Leave for 10 minutes or more after ig-
nition.
If hour counter indicates 200 hours
(USH102D) or 300 hours (HBO103W/2),
replace it.
Center mercury burner.
Connect firmly.
Try perform shutter operation again.
Remove turret and place fluorescence
mirror unit in correct position securely.
Re-attach motorized nosepiece by in-
serting it all the way, until stopped.
Page
10
33
13
–
–
31
30
17