c. Examples:
c-1. 40X image moves to the left, and
6.3X image to the right.
c-2. Lower 40X drum lens until 40X image
is aligned with horizontal center
line of 16X.
c-3. Move 40X drum lens to the right
until 40X image is at equidistant
from north-south optical axis of 16X.
c-4. Move B lens to the right until image
of 40X and 6.3X are coincided with
image of 16X.
5-3) Adjustment of image movement of 25X and
lOX in north-south and east-west direc-
tions.
Coincide optical axes of 25X
and lOX to that of 16X by movement of
25X drum lens.
5-4) Steps 5-2) and -3) sometimes require
repetition until adjustment is complete.
Ascertain coincidence of optical axes
of 40X, 25X, 16X, lOX and 6.3X.
6.
Adjustment of Left Diopter
6-1) Insert eyepiece GlOX into right eye-
piece tube, bring drum lens 16X into
path, and focus on specimen by use of
focusing telescope.
6-2) Coincide focus of left drum lens 16X
to
that of right 16X.
a. When focus is deviated
to
the
"+"
side (or eyepiece tube is length-
ened), move prism P2 backward until
focus is obtained by aligning
helicoid ring
to
position "O".
- 15 -
c-1.
c-2.
4-0X
c-3.
t:f'
c-4.
5-3)
There is correspondence between
the image movements of 25X and 40X;
lOX and 6.3X, respectively.
This adjustment does not require
movement of B lens.
6-1)
Make it sure that when helicoid
ring on left eyepiece tube is
positioned
to
"O", left drum lens
16X must be brought into focus.
a.