Olympus CelonLab ENT 지침 - 페이지 8
{카테고리_이름} Olympus CelonLab ENT에 대한 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus CelonLab ENT 10 페이지. Safety check
6. Earth resistance (according to IEC 60601-1 and IEC 62353)
No.
Test equipment:
Test accessories: Power cord
Connect the unit with an electrical safety tester according to the tester's instructions for use and measure
360
the following parameters against metal parts which can be touched:
370
Protective earth resistance:
Documentation
Document the test results in "SCR.991007".
380
7. Current and power consumption
No.
Test equipment:
Test accessories: Power cord
Connect the unit with an electrical safety tester according to the tester's instructions for use. Connect
390
HF-output 1 and 2 of the unit (see fig. 3) via the connection cables with the 100 Ω load resistor. Set the
power level of the unit to "25" and press the footswitch to measure the following parameters:
Attention: Start measuring, if not already done, after setting the power level to "25"and make
sure to release the footswitch before starting with a new measurement.
400
Current consumption under load at 100...120 V~:
410
Power consumption under load at 100...120 V~:
Documentation
Document the test results in "SCR.991007".
420
© Celon AG
,
Issue date: 01.09.2008
SCI.991007-1.4
Safety Check
Instruction.doc
Safety Check Instruction
Electrical safety tester (Example: Unimet 1000 ST, Bender)
Test probe
Footswitch
Electrical safety tester (Example: Unimet 1000 ST, Bender)
Load resistor 100 Ω, 25 W (or more), 5 % tolerance, low inductive part, short time
load
Connection cables, HF-output (banana) to the load resistor
Footswitch
Procedure
≤ 0.3 Ω
Procedure
1.0 A
I
L
110 VA
S
L
Page 8 / 9