AccuLine 40-6540 서비스 매뉴얼 - 페이지 12
{카테고리_이름} AccuLine 40-6540에 대한 서비스 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. AccuLine 40-6540 13 페이지. Automatic rotating laser
4.2.1. Z-axis Calibration
1.
Place the unit into calibration mode as discussed above.
2.
Position the unit so that the Z axis directing to the two targets.
3.
Press "X/Y" key on the remote control to toggle calibration control to the X-axis
4.
Press the "UP Arrow" and "Down Arrow" keys, respectively to adjust the laser height to coincide with
the zero position of the target
5.
Press the "Enter" key on the remote control to accept the calibration value, noting that the status indicator
goes "off"
6.
After calibration is completed, make sure to power off the unit and then power on again to activate the
calibration.
4.3. Calibration for sound alarming range
If the alarm range is obviously biased, one side exceeds 4.5 degree and other side is much less than 4.5 degree, it needs the
calibration.
Calibration method is as following:
a) Remove the top cover and the shell module
b) Put the unit on the leveling platform, turn on the unit to have it self-leveled and turn it off then.
c) Adjust the touching staff to the circle center of the touching ring on the alarming part.
d) Assemble the housing and the top cover on.
Prepared By: Tim Wojo
Revision - Date: 0 – 20080430
40-6540 Service Manual
Touching point of alarming
part
Page 12 of 13