Olympus CIX100 사용 설명서 - 페이지 45
{카테고리_이름} Olympus CIX100에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus CIX100 46 페이지. Cleanliness inspector system
Olympus CIX100에 대해서도 마찬가지입니다: 시스템 설정 매뉴얼 (38 페이지), 설치 매뉴얼 (14 페이지)
Sample Holders
9
Exchanging the stage insert of a CIX100 system for
metallurgical applications
When wanting to inspect metallurgical samples, it may be sensible to exchange the
stage insert with a stage insert without openings for sample holders. When wanting
to remove the stage insert, follow all instructions given in the stage manual.
Push the stage insert to the front left (towards the red dot marking) and
subsequently lift the stage insert starting with the right back edge.
Be careful not to affect the clamps fixing the stage insert.
When placing samples for metallurgy on the stage, keep in mind that the Z-range
might not be suitable for the sample height. Carefully follow all safety instructions.
Detailed information regarding the stage insert can be found in the manual
delivered by the stage insert itself.
45