Attaching the Breath Shield
11
(MX-BSH or MX-BSH-ESD-2)
@
1. Attach the two breath shield mounting brackets @ on the breath shield
² using the provided screws (shorter ones).
For the present, tighten the screws temporarily using the Allen screw-
driver. (If they are attached completely, it becomes impossible to attach
the breath shield to the microscope frame.)
2. Align the attaching holes | on the mounting brackets with the breath
shield screw holes ³ on the microscope, and clamp using the provided
screws (longer ones).
3. Now tighten the screws that have been secured temporarily in step 1
above.
# Tighten the screws lightly to prevent the breath shield from cracking.