Olympus U-UCD8A 사용 설명서 - 페이지 23
{카테고리_이름} Olympus U-UCD8A에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus U-UCD8A 40 페이지. Motorized universal condenser
Problem
c. Image glares and resolution
is low under brightfield obser-
vation.
d. Ring slit does not align with
phase plate of objective.
e. Darkfield contrast perfor-
mance is inadequate.
f. Polarizing performance is in-
sufficient.
Cause
Aperture iris diaphragm is stopped down
excessively.
Top lens is swung out.
Incorrect optical element is inserted in
light path.
Incorrect objective is inserted in light path.
Top lens is swung out.
Aperture iris diaphragm is stopped down.
Incorrect optical element is inserted in
light path.
Incorrect objective is used.
Darkfield ring is not centered correctly.
Polarizer is not inserted in light path.
Analyzer is not inserted in light path.
Optical element is engaged in light path.
Aperture iris diaphragm is opened.
Remedy
Open aperture iris diaphragm to proper
diameter.
Swing top lens into light path when 10X
to 100X objectives are in use.
Engage optical element that matches
objective in use by rotating turret.
Engage correct objective into light path.
Swing top lens in.
Open aperture iris diaphragm.
Engage darkfield ring in light path.
See tables in "Optical Elements and
Compatible Objectives" on page 25.
Center darkfield ring correctly.
Engage polarizer in light path.
Engage analyzer in light path.
Engage empty position by rotating turret.
Either stop down aperture iris diaphragm.
U-UCD8A
Page
8
–
25
25
–
16
25
25
16
18
17
–
–
20