AccuMed AT2102 사용자 설명서 - 페이지 11
{카테고리_이름} AccuMed AT2102에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. AccuMed AT2102 13 페이지. Infared thermometer
AccuMed AT2102에 대해서도 마찬가지입니다: 빠른 시작 매뉴얼 (6 페이지)
AT2102.
Never
batteries. Use only batteries for single
usage.
IX. TECHNICAL SPECIFICATIONS
1. Normal using condition
Ambient temperature: 10°C ~ 40°C (50°F ~ 104°F)
Relative humidity: ≤85%
2. Batteries: DC 3V (2 x AAA batteries)
3. Unit size: 130 x 45 x 55 mm (L x W x H)
4. Unit weight (without batteries): 75g
5. Temperature display resolution: 0.1°C (0.1°F)
6. Measuring range:
In body mode: 32°C ~ 42.9°C (89.6°F ~ 109.2°F)
In surface temp mode: 0°C ~ 60°C (32°F ~ 140°F)
In room mode: 0°C ~ 40°C (32°F ~ 104°F)
7. Precision:
32.0°C ~ 35.9°C (89.6°F ~ 96.6°F)
36.0°C ~ 39.0°C (96.8°F ~ 102.2°F)
39.1°C ~ 42.9°C (102.4°F ~ 109.2°F)
8. Consumption: ≤150mW
9. Accuracy: ± 0.3°C (0.6°F)
10. Measuring distance: 3cm ~ 5cm (1.2in ~ 2in)
11. Automatic power-off: 30 sec
12. Memory:32 sets
※ Note: The Infrared Non-contact Thermometer Model AT2102 can take
temperature readings below 32°C or above 42.9°C (89.6°F to 109.2°F) but
precision is not guaranteed outside of this range.
LONGEVITY OF THE PRODUCT
The AT2102 was developed for intense and professional use. Its longevity is
guaranteed for 40,000 readings.
X. MAINTENANCE OF THE PRODUCT
- The protective glass over the lens is the most important and fragile part of the
thermometer. Please take great care of it.
- Clean the glass with cotton fabric, wet with 95% alcohol.
- Do not use other batteries than mentioned batteries. Do not recharge non-
use
rechargeable
±0.3°C (±0.6°F)
±0.2°C (±0.4°F)
±0.3°C (±0.6°F)
9