VIII. SETTING AND FUNCTION OF MENU
1. Switch on the device:
Press the "On/Scan" button for one second, after the screen panel is in full display, it will
enter the standby mode with the sign "---°C" or "----°F".
Then press the "On/Scan" button again, you will get the measuring result in 1 second. But if
there is no more operation, it will turn off in 30 seconds automatically.
2. Setting the mode when the device is On:
a. Press "MODE" button, and the screen will display: Body...°C
b. Press again "MODE" button and the screen will display: Room...°C
c. Press again "MODE" button and the screen will display: Surface
Temp...°C
Note: The thermometer default is set to BODY mode.
3. Switching between Celsius and Fahrenheit:
Press the "°C/ °F" button to transfer between Celsius and Fahrenheit.
4. Reviewing Stored Memory:
Press the "MEM"(Memory) button, which will then display the last temperature, and allows for
a view of the last 32 measurements.
Important!
The surface temperature differs from the internal body temperature. To obtain the internal
temperature, always use the "BODY" mode. Please make sure to select "SURFACE TEMP"mode
for an external area reading.
5. When the device is on, press "LANGUAGE" button to enable/disable the voice.
When the screen shows "ON", the voice enabled.
When the screen shows "OFF", the voice disabled.
6. Recalibration of device via the "F-4" MENU
When there is the difference between AT2103 and a mercury thermometer, and you believe the
mercury thermometer is more accurate, you can use the recalibration function to adjust the
AT2103 to make it the same as a mercury thermometer.
8